Mostrar el registro sencillo del ítem

[en] DEVELOPMENT OF METHODOLOGY FOR CALIBRATING PIEZOELECTRIC TRANSDUCERS USED IN INTERNAL BALLISTICS FOR TRANSIENT PRESSURE MEASUREMENTS

dc.contributorALCIR DE FARO ORLANDO
dc.contributorALCIR DE FARO ORLANDO
dc.creatorALFREDO JORGE BAHIA HEINE
dc.date2004-12-30
dc.date.accessioned2022-09-21T21:44:15Z
dc.date.available2022-09-21T21:44:15Z
dc.identifierhttps://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=5851@1
dc.identifierhttps://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=5851@2
dc.identifierhttp://doi.org/10.17771/PUCRio.acad.5851
dc.identifier.urihttps://hdl.handle.net/20.500.12032/43025
dc.description[pt] A presente pesquisa de mestrado, de interesse da indústria de armamento, mais especificamente no controle de qualidade de lotes de munição, relaciona-se com o desenvolvimento de metodologia para calibração de transdutor piezelétrico para medição de pressão transiente cuja duração é da ordem de milisegundos. Como sua principal contribuição, o trabalho disponibiliza metodologia própria de calibração ainda não disponível no País, atribuindo confiabilidade a medições realizadas em fenômenos associados ao que se denomina balística interna, que se constitui no estudo científico dos processos que ocorrem no interior de uma arma de fogo a partir do instante de iniciação (ignição) do propelente. Não obstante a faixa de pressão estudada, de 34,7 a 349,8 bar), ser baixa se comparada com as pressões desenvolvidas na balística interna, que podem até superar 6000 bar, a simples substituição do gerador de pressão por outro de maior capacidade é suficiente para a calibração numa faixa mais larga utilizando-se a mesma metodologia. O complexo processo de balística interna foi simulado em condições laboratoriais pela submissão brusca do sensor piezelétrico, objeto da calibração, originalmente montado em uma câmara diminuta (cerca de dez vezes menor que o volume total de fluido pressurizado) de óleo à pressão atmosférica, a um pico de pressão induzido pela comunicação da câmara a uma linha de óleo pressurizada à pressão de calibração pré-determinada através de uma bomba hidráulica, pressão esta que caracteriza um ponto de calibração. Complementando o aparatus experimental utilizado para simular o fenômeno da balística interna, foi utilizado um sistema de medição com tempo de resposta compatível (rise time < 5 us) à ordem de grandeza da duração do fenômeno estudado. O aparatus experimental utilizado é tal que permite alterar a duração desses eventos por meio de uma válvula de restrição, o que resultou em eventos com duração de 2 a 2200 ms. O monitoramento do sinal de resposta do transdutor piezelétrico no curso da calibração realizada permite estabelecer o tempo-limite associado ao qual o sinal mantém-se isento de indesejável atenuação, considerada fonte de erro na medição de pressão. O processo de calibração desenvolveu-se para 10 pontos distintos, perfazendo um total de 340 experimentos (variando-se o transiente de pressão imposto ao transdutor), monitorando-se o sinal de resposta do transdutor contra o valor da pressão medida por um manômetro calibrado pelo laboratório de pressão da PUC-Rio, credenciado pelo INMETRO e, portanto, assegurando rastreabilidade das medições ao Sistema Internacional de Unidades. Alterações no aparatus experimental permitem que calibrações usualmente realizadas em regime de pressurização sejam também realizadas em regime de despressurização, viabilizando, dentre outras análises possíveis, o estudo de estabilidade e histerese do sinal do transdutor. O método experimental desenvolvido não apenas mostrou-se tecnicamente viável e adequado às condições laboratoriais da faixa de pressão na qual foi validado (34,7 a 349,8 bar), como também se mostrou viável economicamente. A incerteza associada a esse método, obviamente, é maior que aquela obtida com uma balança de peso morto, que se baseia num método primário. A incerteza associada ao método de calibração desenvolvido no presente trabalho de pesquisa pode ser diminuída se for usado um sistema de medição de tensão com melhor exatidão.
dc.description[en] The current research, of armaments industry s interest, more specifically in the quality control of ammunition lots, is related to the development of methodology for calibrating piezoelectric transducers for transient pressure measurements associated with events whose duration is of the order of milliseconds. As its main contribution, this work presents proper methodology, not available yet in the Country, to ensure reliability to measurements of pressure in internal ballistics, which is the scientific study of the operating processes within the gun from at the moment when the burning process of the propellant is initiated. In spite of the low transient pressure range studied, 34,7 to 349,8 bar, when compared with the pressures developed in internal ballistics, up to 6000 bar, the simple replacement of the pressure generator for another one of larger capacity is enough to allow for calibration in a wider pressure range using the same methodology. The complex process associated to internal ballistics was simulated in laboratorial conditions by the abrupt exposure of the piezoelectric sensor, object of the calibration, planned and originally mounted in a small oil chamber (about ten times smaller than the total volume of fluid pressurized) maintained at atmospheric pressure, to a pressure peak introduced by the communication of the small chamber to a line of oil pressurized at a desired calibration pressure level by means of a hydraulical pump used to feed the oil system. Each calibration pressure condition corresponds to what is called a calibration point. Complementing the experimental apparatus used to simulate the phenomenon of the internal ballistics, an adequate measurement system (rise time < 5 us, therefore adequate to the time duration of the studied phenomenon) was used. The experimental apparatus employed is such that the time duration of the events could be changed by means of a restriction valve, allowing events with duration from about 2 to 2200 ms. Within the calibration process, the monitoring of the piezoelectric transducer output signal allowed for establishing a specific time-limit to ensure that the signal remained exempt of undesirable attenuation, considered a source of error in the measurement of the transient pressure. The calibration process was developed for 10 distinct points, totalizing 340 experiments (varying the transient pressure to which the transducer is exposed), monitoring the output signal of the transducer against the value of the pressure measured with a manometer calibrated in the accredited pressure laboratory of PUC-Rio, traceable to the national standard maintained by the Brazilian National Metrology Laboratory (INMETRO) and, therefore, ensuring overall traceability of measurements to the International System of Units (SI). Modifications introduced in the experimental apparatus used allow that calibrations usually carried out under a regime of pressurization may also be carried out under a regime of unpressurization, making possible other studies such as stability and hysteresis of the transducer s output signal. From the technical point of view the experimental method proved to be feasible and suitable to the laboratorial conditions for the pressure range to which it was validated (34,7 to 349,8 bar). It also revealed economically viable. Although validated, the uncertainty associated with the proposed method, obviously is greater than the uncertainty associated with a dead weight tester, which is based on a primary method. However, the uncertainty associated with the method of calibration herein proposed can be minimized whenever a more accurate voltage measurement system is used.
dc.languagept
dc.publisherMAXWELL
dc.subject[pt] METROLOGIA
dc.subject[pt] BALISTICA INTERNA
dc.subject[pt] TRANSDUTOR
dc.subject[pt] INCERTEZA DE MEDICAO
dc.subject[pt] PIEZELETRICO
dc.subject[pt] CONTROLE DE QUALIDADE
dc.subject[pt] CALIBRACAO
dc.subject[en] METROLOGY
dc.subject[en] INTERNAL BALLISTICS
dc.subject[en] TRANSDUCER
dc.subject[en] UNCERTAINTY OF OF MEASUREMENT
dc.subject[en] PIEZOELECTRIC
dc.subject[en] QUALITY CONTROL
dc.subject[en] CALIBRATION
dc.title[pt] DESENVOLVIMENTO DE METODOLOGIA DE CALIBRAÇÃO DE TRANSDUTOR PIEZELÉTRICO DE PRESSÃO UTILIZADO EM BALÍSTICA INTERNA
dc.title[en] DEVELOPMENT OF METHODOLOGY FOR CALIBRATING PIEZOELECTRIC TRANSDUCERS USED IN INTERNAL BALLISTICS FOR TRANSIENT PRESSURE MEASUREMENTS
dc.typeTEXTO


Ficheros en el ítem

FicherosTamañoFormatoVer

No hay ficheros asociados a este ítem.

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem


© AUSJAL 2022

Asociación de Universidades Confiadas a la Compañía de Jesús en América Latina, AUSJAL
Av. Santa Teresa de Jesús Edif. Cerpe, Piso 2, Oficina AUSJAL Urb.
La Castellana, Chacao (1060) Caracas - Venezuela
Tel/Fax (+58-212)-266-13-41 /(+58-212)-266-85-62

Nuestras redes sociales

facebook Facebook

twitter Twitter

youtube Youtube

Asociaciones Jesuitas en el mundo
Ausjal en el mundo AJCU AUSJAL JESAM JCEP JCS JCAP