Description
Objectives: The accuracy of cone-beam technique, cone-beam computed tomography (CBCT), compared with that of the multislice spiral CT (MSCT), for image-based linear measurements of midpalatal suture was assessed. Material and Methods: Two measurements were performed by one investigator on the dry skull by using one digital caliper and in the axial cuts by using software. A 2D object-based image registration process was applied to determine the best affine transformation that maps a 2D input image (CBCT) in a reference image (MSCT). Results: The value of the intraclass correlation coefficient was approximately 0.9%. The results suggest that differences between two scanners did not exist (observation 1, p = 0.964 and observation 2, p = 0.795).With regard to the dry skull and the image, the significance probabilities equaled zero (observation 1, p = 0.002 and observation 2, p = 0.004), therefore, indicating significant differences. Conclusion: Measurements acquired in the images were similar and these findings contribute to stimulate the use of CBCT for evaluation of the maxillary expansion procedure. © 2011 Taylor & Francis.